The AirSense 10 AutoSet is a premium auto-adjusting therapy device. It features integrated humidifier with ClimateLineAir™ for climate control, AutoRamp™ with sleep onset detection, expiratory pressure relief (EPR™) and Easy-Breathe technology. The AirSense 10 AutoSet provides advanced event detection, including indications of central sleep apnoea and Cheyne-Stokes respiration. Optional pulse oximetry monitoring is also available. Modes include AutoSet and CPAP. With the AirSense 10 AutoSet patient trial devices you can be more confident than ever before during their critical early days of adjusting to treatment.
ResMed’s AutoSet algorithm now allows for additional customisation of therapy with the new AutoSet Response feature which offers all the benefits of the enhanced AutoSet algorithm with gentler pressure increases for additional comfort.
The AirSense 10 AutoSet differentiates between obstructive and central apnoeas and can also detect indications of Cheyne-Stokes respiration, so you can be confident that your patients are receiving appropriate therapy.
AirSense 10 AutoSet | |
---|---|
CPAP | ✓ |
AutoSet | ✓ |
AutoSet for Her | |
Cheyne-Stokes respiration detection | ✓ |
Central apnea detection | ✓ |
RERA reporting | |
AutoSet algorithm with AutoSet Response |
✓ |
AutoRamp with sleep onset detection | ✓ |
Climate Control | ✓ |
Enhanced Climate Control Auto |
✓ |
Advanced HumidAir technology | ✓ |
Expiratory Pressure Relief (EPR™) | ✓ |
SmartStart™ | ✓ |
Dimensions (H x W x D) (Device and water tub) | 4.57" x 10.04" x 5.91" | |
---|---|---|
Weight | 44 oz | |
Air filter | Standard: Average arrestance: > 75% for ~7 micron dust |
Hypoallergenic: Efficiency: >98% for ~7-8 micron dust. >80% for ~0.5 micron dust. |
Aircraft use | Meets Federal Aviation Administration requirements for all phases of air travel. | |
Operating pressure range | 4 to 20 cm H2O |
Pressure level measured according to ISO 17510-1:2007 (CPAP mode): |
---|
SlimLine or Standard air tubing | 26.6 dBA with uncertainty of 2 dBA |
Power level measured according to ISO 17510-1:2007 (CPAP mode): |
---|
SlimLine or Standard air tubing | 34.6 dBA with uncertainty of 2 dBA |
Sound (Declared dual-number noise emission values in accordance with ISO 4871:1996.) |
Air tubing | ClimateLineTM Air | SlimLineTM | Standard |
---|---|---|---|
Length | 6'6” | 6’ | 6’.6” |
Inner diameter | 0.6" | 0.6" | 0.75" |
Electromagnetic compatibility | Complies with all EMC requirements according to IEC60601-1-2:2007 for residential, commercial and light industry environments. It is recommended that mobile communication devices be kept at least 1m away. |
Kline LR, Carlson P. NCPAP acceptance and compliance is altered by humidification. Sleep 1999;22: Suppl. S230